کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
3142334 | 1196782 | 2015 | 7 صفحه PDF | دانلود رایگان |
ObjectiveTo evaluate the accuracy of craniomaxillofacial resections performed with an image-guided surgical sagittal saw.Material and methodsTwenty-four craniomaxillofacial resections were performed using an image-guided sagittal saw. Surgical outcomes were compared with a preoperative virtual plan in terms of the resected bone volume, control point position and osteotomy trajectory angle. Each measurement was performed twice by two independent observers.ResultsThe best convergence between the planned and actual bone resection was observed for the orbital region (6.33 ± 4.04%). The smallest mean difference between the preoperative and postoperative control point positions (2.00 ± 0.66 mm) and the lowest mean angular deviation between the virtual and actual osteotomy (5.49 ± 3.17 degrees) were documented for the maxillary region. When all the performed procedures were analyzed together, mean difference between the planned and actual bone resection volumes was 9.48 ± 4.91%, mean difference between the preoperative and postoperative control point positions amounted to 2.59 ± 1.41 mm, and mean angular deviation between the planned and actual osteotomy trajectory equaled 8.21 ± 5.69 degrees.ConclusionThe results of this study are encouraging but not fully satisfactory. If further improved, the hereby presented navigation technique may become a valuable supporting method for craniomaxillofacial resections.
Journal: Journal of Cranio-Maxillofacial Surgery - Volume 43, Issue 7, September 2015, Pages 1021–1027