کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3143208 1196809 2010 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Pullout strength of a biodegradable free form osteosynthesis plate
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Pullout strength of a biodegradable free form osteosynthesis plate
چکیده انگلیسی

SummaryThe Inion® Free Form Plate is a newly designed biodegradable plate. After drilling through the plate and tapping, a biodegradable screw can be inserted, followed by removal of the screw head. As an alternative a countersink screw can be used.Aim of the study was to compare the mechanical properties of the 1.4 mm Free Form Plate with the 2.0 mm conventional shaped plate.Mechanical testing of the plate pullout strength was conducted for the Inion® Free Form Plate fixed with an Inion OTPS™ 2.0 × 20 mm Screw. In addition, the failure mode was reported.Overlapping confidence levels were found with regard to the yield load, first peak load and maximum load, when comparing the Free Form Plate and the conventional 4-hole plate. The Free Form Plate fixed with a screw with head and countersink showed the highest stability at maximum load. The results of the mechanical stability testing showed no significant differences between the tested plates. The main failure mode was a failure of the screw shaft.The results of the current investigation imply that the 1.4 mm Free Form Plate could be used as an alternative to the 2.0 mm conventional shaped plate.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Cranio-Maxillofacial Surgery - Volume 38, Issue 7, October 2010, Pages 517–521
نویسندگان
, , , , , , ,