کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3154583 1198051 2008 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Strength Analysis of 6 Resorbable Implant Systems: Does Heating Affect the Stress-Strain Curve?
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Strength Analysis of 6 Resorbable Implant Systems: Does Heating Affect the Stress-Strain Curve?
چکیده انگلیسی

PurposeThe objective was to directly compare the strength of 6 different resorbable implant plating systems using an in vitro model before and after heating.Materials and MethodsRed oak wood was cut and fixated using various resorbable plates and screws. Vertical load was applied and the specimens fractured, while a test machine gathered data. This was repeated after heating of the specimens.ResultsSeveral parameters were analyzed, and force versus displacement curves were plotted for each specimen.ConclusionsThere were no statistically significant differences for total maximum loads between heat-treated and non heat-treated specimens. There were differences in strengths amongst the various systems.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Oral and Maxillofacial Surgery - Volume 66, Issue 12, December 2008, Pages 2493–2497
نویسندگان
, , , , , ,