Biomechanical Testing of Different Osteosynthesis Systems for Segmental Resection of the Mandible
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Biomechanical Testing of Different Osteosynthesis Systems for Segmental Resection of the Mandible
چکیده انگلیسی
Unilock plates fixed with locking screws have a higher long-term stability than the Reconstruction 2.4 system. A 1-mm gap between the plate and mandible does not lead to early screw failure in the Unilock 2.4 system with locking screws. The Reconstruction 2.7 system seems superior if well contoured, because 2 of those plates reached the runout limit; however, this system is not as easy to handle as the 2.4 systems, and good contouring is difficult to achieve. Therefore, we consider the Unilock 2.4 system with locking screws the best choice.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Oral and Maxillofacial Surgery - Volume 65, Issue 5, May 2007, Pages 924-930