PurposeThe goal was to measure heat generated in bone by 3 implant drill systems after repeated drilling and sterilization.Materials and MethodsTemperature was measured with thermocouple technology in vitro using the bovine femoral cortical bone model. Intermittent drilling was accomplished at a constant 2.4-kg load and drill speed of 2,500 rpm. External irrigation at 40 mL/min with normal saline was used. Three implant drill systems—system A (triple twist drills with a relief angle), system B (triple twist drills without a relief angle), and system C (double twist drills with a relief angle)—were evaluated and heat was measured at the final drill in the drilling sequence (4.0 mm or 4.2 mm) at a depth of 15 mm. Thermocouples were placed 0.5 mm from the osteotomy at a depth of 15 mm. Heat measurements were recorded out to 25 uses.ResultsResults showed temperature increased with multiple uses. System A and C drills had temperature measurements that were below 47°C, even after 25 uses. System B drills had temperatures that exceeded 47°C from the initial use. Light microscopy showed little drill wear even after 25 uses.ConclusionsDrill geometry plays a major role in heat production and may explain the increased temperature readings seen in system B. These drills lack relief angles and have the smallest clearance angles of the 3 systems. It also has fewer drills in its drilling sequence compared with systems A and C. This study shows that temperatures increase when drills are used multiple times. Systems A and C had acceptable temperature measurements out to 25 uses. System B drills showed significantly higher heat production with little visual signs of wear.
Journal: Journal of Oral and Maxillofacial Surgery - Volume 64, Issue 2, February 2006, Pages 265–269