BackgroundThe purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods.MethodsForty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 135° AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD).ResultsResults for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm−1, respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm−1. The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively.ConclusionsThe results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods.
Journal: Injury - Volume 43, Issue 6, June 2012, Pages 802–806