The Malaria Pf Rapid Test Device Acon® (Acon Labs) and the pan HRP2/aldolase RDT, Malaria P.f/Pan Rapid Test Device Acon® (Acon Labs), performances were evaluated for malaria species diagnosis in 592 febrile patients living in Gabon using microscopy as gold standard. Sensitivities were equal or above 96.0% for Plasmodium falciparum detection, of 62.5% for non–P. falciparum malaria species detection and higher in younger children (100%). Negative predictive values were greater than 97.0%. Acon®HRP2 had a higher specificity (96.6%) and lower false-positive (FP) rate (9.3%) compared to Acon®Pf/Pan, which had a specificity of 87.3% and a FP rate of 27.1% (P < 0.01). Overall, 32.5% of all Acon® Pf/Pan tests resulted in a “faint band” with only 2 resulted from samples with a parasitemia below 100 p/μL. The accuracy of Acon®HRP2 RDT for the diagnosis of P. falciparum infection is confirmed. However, the high FP rate observed with Acon®Pf/Pan is a limitation for its use.
Journal: Diagnostic Microbiology and Infectious Disease - Volume 77, Issue 1, September 2013, Pages 58–63