Biomechanical and Computed Tomography Analysis of Adjustable Femoral Cortical Fixation Devices for Anterior Cruciate Ligament Reconstruction in a Cadaveric Human Knee Model
ترجمه فارسی عنوان
تجزیه و تحلیل توموگرافی بیومکانیک و محاسبات دستگاه های ثابت کننده فتق ریزی فمورال برای بازسازی مجدد سفتی قدامی در یک مدل کراوات زای انسان
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی

PurposeTo evaluate and compare two adjustable femoral cortical suspensory fixation devices used for anterior cruciate ligament reconstruction through a novel, direct computed tomography (CT) analysis metric and biomechanical laxity testing in a matched cadaveric human knee study.MethodsAnterior cruciate ligament reconstructions with bovine tendon grafts were performed using two adjustable femoral cortical suspensory fixation devices (RigidLoop Adjustable [DePuy Synthes Mitek, Raynham, MA] and TightRope [Arthrex, Naples, FL]) in 12 knees (6 matched pairs). A mechanical testing series was used to determine each knee's laxity in the intact condition. After reconstruction, each specimen was again tested for laxity and also imaged with CT. The laxity testing and CT imaging were then repeated after 1,000 cycles of anteroposterior loading on each knee to compare changes in laxity for the two fixation devices and to visualize changes in button-to-graft distance migration through a three-dimensional CT imaging method.ResultsNo significant differences were found between the two fixation groups' laxity measures after reconstruction (all P values ≥ .620) or after cycling (all P values ≥ .211) at any flexion angle. In addition, no significant differences were found between the two groups regarding button-to-graft distance migration (P = .773; mean, 0.61 ± 0.6 mm [95% confidence interval, −0.1 to 1.3 mm] in RigidLoop Adjustable group and 0.53 ± 0.6 mm [95% confidence interval, −0.1 to 1.2 mm] in TightRope group).ConclusionsThere were no significant differences between the two femoral cortical suspensory adjustable-loop devices regarding laxity outcomes or loop displacement as measured by button-to-graft distance migration.Clinical RelevanceUse of either of the adjustable-loop cortical suspensory devices in our analysis would appear to produce similar, acceptable laxity outcomes and minimal effects in terms of device-related loop displacement.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Arthroscopy: The Journal of Arthroscopic & Related Surgery - Volume 32, Issue 2, February 2016, Pages 253–261
نویسندگان
, , , , , ,