Accuracy Analysis of a Novel Electromagnetic Navigation Procedure Versus a Standard Minimally Invasive Method for Arthroscopically Assisted Acromioclavicular Joint Reconstructions
ترجمه فارسی عنوان
تجزیه و تحلیل دقیق یک روش ناوبری الکترومغناطیسی رمان در مقابل یک روش استاندارد حداقل تهاجمی برای بازسازی مکانیکی آکرومیوکلوویکورس با کمک آرتروسکوپی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی

PurposeThe aim of this cadaveric study was to evaluate the accuracy, feasibility, and operation time of a novel electromagnetic navigation system (ENS) and procedure for transclavicular-transcoracoid tunnel placement compared with a standard minimally invasive (SMI) reconstruction method for minimally invasive arthroscopically assisted anatomic acromioclavicular joint reconstruction.MethodsTen arthroscopically assisted electromagnetic-navigated transclavicular-transcoracoid drilling procedures and 10 SMI procedures were performed on 10 human cadavers using 2 TightRope reconstructions (Arthrex, Naples, FL) for each site. Postoperative computed tomography scans were acquired to determine tunnel placement accuracy. Optimal coracoid tunnel placement was defined according to the anatomic insertions of the trapezoid and conoid ligaments with the oblong button placed in a center-base position at the coracoid undersurface without cortical breach or fracture. Both reconstruction procedures were performed without fluoroscopy.ResultsSuccessful tunnel placement was accomplished in 98.8% using the ENS method and in 83.8% using the SMI procedure (P = .087). The mean overall operation time was 28.5 ± 6.6 minutes for the ENS method and 35.2 ± 3.9 minutes for the SMI method (P = .012). The ENS procedure required no directional readjustments or restarts. Drill misguidance with subsequent cortical breach occurred twice in the SMI group. In both groups no fractures were documented.ConclusionsIn this descriptive laboratory study, both the ENS and the SMI techniques achieved the desired anatomic reconstruction. Compared with the SMI method, the ENS procedure showed higher accuracy, required a shorter operation time, and was associated with no complications.Clinical RelevanceLength of surgery, drilling accuracy, and reduction of radiation exposure in acromioclavicular joint reconstruction procedures affect the safety of both patients and surgeons.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Arthroscopy: The Journal of Arthroscopic & Related Surgery - Volume 30, Issue 8, August 2014, Pages 928–935
نویسندگان
, , , , , , , ,