کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4045776 1603548 2011 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Complication of the Use of a Radiofrequency Device in Arthroscopic Shoulder Surgery: Second-Degree Burn of the Shoulder Girdle
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Complication of the Use of a Radiofrequency Device in Arthroscopic Shoulder Surgery: Second-Degree Burn of the Shoulder Girdle
چکیده انگلیسی

Radiofrequency ablation (RF) energy devices are increasingly being used in arthroscopic surgery. This is a case report of a 53-year-old man who had second-degree burns of the shoulder and chest wall as a result of fluid overheating due to RF. During an extensive arthroscopic bursal dissection of the left shoulder, the suction device on the RF wand and the valve was left open, allowing fluid to drip onto the chest and arm of the patient. Three days after his surgery, on his first postoperative visit, the dressings were removed, and the patient was noted to have severe blistering on the lateral aspect of his arm and on the anterolateral aspect of his left chest wall. One year after the surgery, he still had residual scarring that was only mildly intermittently pruritic. The use of RF during arthroscopic surgery is very useful clinically; however, it is important to have a strict understanding of the potential hazards the thermal energy can cause to the surrounding soft tissue.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Arthroscopy: The Journal of Arthroscopic & Related Surgery - Volume 27, Issue 1, January 2011, Pages 136–141
نویسندگان
, , ,