PurposeThe purpose of this study was to compare the size of the entrance of the notch, as measured arthroscopically (2-dimensionally), with the volume of the notch as measured by use of computed tomography (CT) (3-dimensionally).MethodsFor 20 cadaveric knees, the dimensions of the notch entrance were measured arthroscopically, and the notch volume was measured by use of CT. The correlation between the size of the notch entrance and the notch volume was calculated. Intraobserver reliability and interobserver reliability of the arthroscopic and CT measurements were tested.ResultsThe Pearson correlation coefficients between CT-assessed notch volume and arthroscopically assessed notch height and width at the bottom, middle, and top of the notch were 0.603, 0.506, 0.551, and 0.642, respectively. The intraobserver reliability and interobserver reliability of the arthroscopic measurements were above 0.962 and 0.819, respectively, and 0.983 and 0.975, respectively, for the CT measurements.ConclusionsThere were only moderate correlations between arthroscopic notch measurements and notch volume. Both the arthroscopic and CT measurements proved highly reliable.Clinical RelevanceThe moderate correlation between 2-dimensional and 3-dimensional notch measurements warrants caution concerning the use of either measurement for assessing risk for anterior cruciate ligament injury or as justification for notchplasty until studies between the relation of 3-dimensional notch volume and anterior cruciate ligament injury are conducted.
Journal: Arthroscopy: The Journal of Arthroscopic & Related Surgery - Volume 27, Issue 2, February 2011, Pages 207–212