کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4060138 | 1603985 | 2015 | 11 صفحه PDF | دانلود رایگان |

The articular surface replacement (ASR) total hip arthroplasty (THA) showed accelerated failure rates due to adverse-reaction to metal debris (ARMD). Literature correlating preoperative with intraoperative revision findings respectively post-revision outcome results are rare. 30 of 99 available ASR THA were revised due to ARMD. Mean post-revision follow-up term was 2.3 years. In part, preoperative data did not correlate with intraoperative revision findings. ARMD was even found in asymptomatic patients with non-elevated ion levels. Postoperative pain and metal ions decreased significantly (P≤0.016). Cobalt decreased faster than chrome. Patients with intraoperative pseudotumors, osteolysis or bilateral THA did not have higher pre- or postoperative ion values (P≥0.053). Females showed higher postoperative chrome levels (P=0.031). One major post-revision complication (femoral nerve palsy) and one re-revision (late onset infection) occurred.
Journal: The Journal of Arthroplasty - Volume 30, Issue 9, September 2015, Pages 1607–1617