کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4060345 | 1603994 | 2015 | 6 صفحه PDF | دانلود رایگان |
A prospective clinical study of 50 patients was conducted to validate a new method of imageless computer navigated hip arthroplasty. The new method enables the surgeon to acquire all registration points with the patient positioned and draped in lateral decubitus position. The final component orientation was measured from post-operative CT scans. The mean error in component position was − 1.1° (SD 3.1°) for inclination and 0.9° (SD 4.3°) for anteversion. This compared favourably with the error of − 1.8° (SD 1.8°) for inclination and − 4.8° (SD 2.7°) for anteversion when using the traditional APP registration. Results show that one can expect the acetabular component to be within a safe zone of ± 10° in 99.8% for inclination and 97.7% for anteversion when using the new lateral registration method. Level of Evidence Level II, Prognostic study.
Journal: The Journal of Arthroplasty - Volume 30, Issue 1, January 2015, Pages 55–60