4060489 | 1603990 | 2015 | 5 صفحه PDF | دانلود رایگان |

Modular neck femoral stems have been associated with adverse local tissue reactions (ALTR), leading to a voluntary recall, but these effects have not been well-characterized. A retrospective review of intraoperative findings and cobalt/chromium levels was performed in 103 hips undergoing revision for ALTR. The average preoperative serum cobalt level was 7.6 μg/L (range 1.1–23 μg/L) and chromium level was 1.8 μg/L (range 0.1–6.8 μg/L). Metallic sludge was noted in 100%, synovitis in 98%, pericapsular rind in 82%, and calcar erosion in 85%. An osteotomy was required for removal in 44%. We concluded that revision of modular neck femoral stems is associated with increased preoperative metal ion levels and stem-neck corrosion. Despite advanced stem explantation techniques, osteotomy was frequently required, leading to increased morbidity.
Journal: The Journal of Arthroplasty - Volume 30, Issue 5, May 2015, Pages 822–826