کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4060519 1603998 2014 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Tibial Plateau Coverage in UKA: A Comparison of Patient Specific and Off-The-Shelf Implants
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Tibial Plateau Coverage in UKA: A Comparison of Patient Specific and Off-The-Shelf Implants
چکیده انگلیسی

Poor tibial component fit can lead to issues including pain, loosening and subsidence. Morphometric data, from 30 patients undergoing UKA were utilized; comparing size, match and fit between patient-specific and off-the-shelf implants. CT images were prospectively obtained and implants modeled in CAD, utilizing sizing templates with off-the-shelf and CAD designs with patient-specific implants. Virtual surgery was performed, maximizing tibial plateau coverage while minimizing implant overhang. Each implant evaluated to examine tibial fit. Patient-specific implants provided significantly greater cortical rim surface area coverage versus off-the-shelf implants: 77% v. 43% medially and 60% v. 37% laterally. Significantly less cortical rim overhang and undercoverage were observed with patient-specific implants. Patient-specific implants provide superior cortical bone coverage and fit while minimizing overhang and undercoverage seen in off-the-shelf implants.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 29, Issue 9, September 2014, Pages 1694–1698
نویسندگان
, , , ,