کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4060710 1604004 2014 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Defining the Errors in the Registration Process During Imageless Computer Navigation in Total Knee Arthroplasty: A Cadaveric Study
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Defining the Errors in the Registration Process During Imageless Computer Navigation in Total Knee Arthroplasty: A Cadaveric Study
چکیده انگلیسی

Computer assisted arthroplasty was introduced as a means to optimally align implants in order to improve function and longevity. The error during the manual registration of landmarks and its effect on component alignment was investigated in this study. Five fresh frozen lower limbs were used and the registration process was performed five times by five surgeons. The error range of the mechanical axis of the femur in the coronal plane was 5.2 degrees of valgus to 2.9 degrees of varus whilst the transepicondylar axis error was 11.1 degrees of external to 6.3 of internal rotation. Those figures suggest that the registration error alone can have a significant effect on the alignment of the implant.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 29, Issue 4, April 2014, Pages 698–701
نویسندگان
, , , , , ,