کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4061139 1604010 2013 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Fixation of Non-Cemented Total Hip Arthroplasty Femoral Components in a Simulated Proximal Bone Defect Model
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Fixation of Non-Cemented Total Hip Arthroplasty Femoral Components in a Simulated Proximal Bone Defect Model
چکیده انگلیسی

An accelerated sequential proximal femoral bone loss model was used to measure the initial stability of three noncemented femoral stem designs: fully porous-coated, proximally porous-coated, and dual-tapered, diaphyseal press-fit (N = 18). Only dual-tapered, diaphyseal press-fit stems remained stable with as much as 105 mm of bone loss, with average cyclic micromotion remaining below 25 μm in ML and below 10 μm in AP planes. In contrast, with proximally coated and fully coated stem designs with circular or oval cross-sections, 60 mm of bone loss, resulting in lower than 10 cm of diaphyseal bone contact length, led to gross instability, increasing average cyclic micromotions to greater than 100 μm prior to failure. Therefore, the results provide support for using a dual-tapered stem in revision cases with proximal bone loss.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 28, Issue 9, October 2013, Pages 1618–1624
نویسندگان
, , , , , ,