کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4061294 1604027 2012 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Early Failure of a Polyethylene Acetabular Liner Cemented Into a Metal Cup
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Early Failure of a Polyethylene Acetabular Liner Cemented Into a Metal Cup
چکیده انگلیسی
In 2002, a patient underwent revision total hip arthroplasty for polyethylene wear. The acetabular cup was well fixed, and it was decided to cement a new polyethylene liner into the existing cup. In 2006, the patient presented with inability to weight bear and easy subluxation of the hip. Revision surgery was performed, and all components were examined postoperatively. Investigation with microcomputed tomography revealed that the liner had plastically deformed at the superior pole resulting in the hip instability. The reasons for this are suspected to be related to the abducted nature of the original cup and an uneven cement mantle. This report suggests that cementation of polyethylene liners into metal cups has limitations and is not appropriate in all circumstances.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 5, May 2012, Pages 820.e5-820.e8
نویسندگان
, , , , , , ,