کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4061805 1604021 2012 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
What are the Predictors for Bead Shedding in Porous-Coated Hip and Knee Arthroplasties? : An Implant Retrieval Study
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
What are the Predictors for Bead Shedding in Porous-Coated Hip and Knee Arthroplasties? : An Implant Retrieval Study
چکیده انگلیسی

Bead shedding has been reported in both hip and knee arthroplasties and can result in increased polyethylene wear and metal ion release. The rates of bead loss were recorded for each device type, and the interfacial bead-substrate shear load to failure and contact areas of the beads were determined. The 3 components with the highest rate of bead loss, namely, S-Rom, Omnifit, and Optifix acetabular shells, recorded the lowest shear loads to failure. Percentage bead-substrate contact area ranged between 7% and 37% and was the most important factor affecting bead shedding. Other factors influencing bead loss in decreasing order of importance were bead fusion, bead size, component type (acetabular or knee), component loosening, and time in situ.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 10, December 2012, Pages 1789–1795