کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4061845 1604020 2013 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Squeaking Hip Arthroplasties : A Tribological Phenomenon
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Squeaking Hip Arthroplasties : A Tribological Phenomenon
چکیده انگلیسی

The clinical incidence of squeaking has been reported with increasing frequency, with ceramic-on-ceramic bearings seemingly most affected. This study investigated potential causes of squeaking in hard-on-hard hip bearings through 2 sets of experimental conditions. Bearing clearance appeared to affect the incidence of squeaking in metal-on-metal surface arthroplasties. The addition of third-body particles to the interface for total hip arthroplasties also affected the incidence of squeaking. In both studies, the incidence of squeaking correlated well with elevated friction. The findings of this study suggest that a likely cause of squeaking in the hip arthroplasty is adverse tribological conditions caused by suboptimal lubrication. There are numerous factors that may cause the suboptimal lubrication, and therefore, it is unlikely that an individual cause for squeaking will be identified.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 28, Issue 1, January 2013, Pages 90–97
نویسندگان
, , , , ,