Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty : A Retrieval Study Using 3-Dimensional Laser Scanning
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty : A Retrieval Study Using 3-Dimensional Laser Scanning
چکیده انگلیسی

The purpose of this study was to evaluate the reliability of the PowerPoint (PP) (2007 Version; Microsoft, Redmond, Wash) method for measuring polyethylene liner wear after total hip arthroplasty. Seventeen retrieved polyethylene liners were included in this study. Wear volumes were calculated using the PP, the Dorr and Wan, and 3-dimensional (3D) laser scanning methods. Spearman correlation coefficients for wear volume results indicated strong correlations between the PP and 3D laser scanning methods (range, 0.89-0.93). On the other hand, Spearman correlation analysis revealed only moderate correlations between the Dorr and Wan and 3D laser scanning methods (range, 0.67-0.77). The PP method can be used to monitor linear wear after total hip arthroplasty and could serve as an alternative method when computerized methods are not available.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 8, September 2012, Pages 1530–1537
نویسندگان
, , , , ,