کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4061971 1604022 2012 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Highly Cross-Linked vs Conventional Polyethylene: No Differences in Rim Notching From Micromotion on Retrieved Acetabular Liners
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Highly Cross-Linked vs Conventional Polyethylene: No Differences in Rim Notching From Micromotion on Retrieved Acetabular Liners
چکیده انگلیسی
Previous literature suggested highly cross-linked acetabular liners demonstrated notching patterns that may be a point of crack origin and rim failure. We examined (1) whether notching patterns and rim cracks existed and demonstrated similar morphological properties in retrieved highly cross-linked and non-cross-linked liners and (2) whether the dimensions of these notches correlated with their duration of implantation. We retrieved a series of 14 identical liners out to an average of 2.03 years. Liners were microscopically examined and then scanned using microcomputed tomography. All liners demonstrated identical notching patterns. Microcomputed tomographic scans demonstrated no signs of crack initiation or rim failure but were able to accurately quantify the notch dimensions. The notching patterns were likely caused by liner-cup micromotion and are better characterized as creep deformation because they did not progress markedly over longer durations of implantation.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 9, October 2012, Pages 1616-1621.e1
نویسندگان
, , , , , , , , , , ,