کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4062400 1604030 2012 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Interface Micromotion of Uncemented Femoral Components from Postmortem Retrieved Total Hip Replacements
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Interface Micromotion of Uncemented Femoral Components from Postmortem Retrieved Total Hip Replacements
چکیده انگلیسی
Axial torsional loads representative of gait and stair climbing conditions were applied to transverse sections of 8 uncemented postmortem retrievals and a high-resolution imaging system with digital image correlation was used to measure local micromotion along the bone-implant interface. For 7 components that were radiographically stable, there was limited micromotion for gait loading (1.42 ± 1.33 μm) that increased significantly (P = .0032) for stair climb loading (7.32 ± 9.96 μm). A radiographically loose component had motions on the order of 2.3 mm with gait loading. There was a strong inverse relationship between the amount of bone-implant contact (contact fraction) (P = .001) and micromotion. The uncemented components had greater contact fraction (41.8% ± 14.4% vs 11.5% ± 10.2%, P = .0033) and less median micromotion (0.81 ± 0.79 μm vs 28.8 ± 51.1 μm) compared to a previously reported study of cemented retrievals.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 2, February 2012, Pages 238-245.e1