کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4062415 1604030 2012 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Fracture of the Outer Metallic Head of the Bipolar Hip Prosthesis: An Unusual Bearing Surface Failure
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Fracture of the Outer Metallic Head of the Bipolar Hip Prosthesis: An Unusual Bearing Surface Failure
چکیده انگلیسی
Fracture of the bearing surface is an infrequent cause of failure of a hip arthroplasty. Although well documented with ceramic heads, fracture of the metallic head is much rarer. We report a case of a fracture of the outer metallic head of a modular cemented bipolar hemiarthroplasty 2 years after the index procedure. Over time, the outer head lost its intended motion and assumed a vertical position. We hypothesized that this position caused asymmetrical loading with stress concentration at the poles, compounded by repeated impingement between the skirted inner cobalt-chromium (Cr-Co) head and the outer stainless steel head of this particular prosthesis. These were supported by the finite element studies. In addition, scanning electron microscopy and energy dispersive x-ray studies showed metallurgical defects that seemed to have initiated and/or accelerated the fracture. Although rare, this mode of failure calls for increased awareness, periodic follow-up, and quality control.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 2, February 2012, Pages 323.e9-323.e12
نویسندگان
, , , , , , , , , ,