کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4062490 1604047 2010 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Assessment of Accuracy and Reliability in Preoperative Templating for Hip Resurfacing Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Assessment of Accuracy and Reliability in Preoperative Templating for Hip Resurfacing Arthroplasty
چکیده انگلیسی

The current study investigated the accuracy and reliability of hip resurfacing component selection based on digital preoperative templating. Four surgeons made a template of preoperative radiographs on 2 occasions for acetabular and femoral components in 50 randomly selected hip resurfacing patients. Component selection reliability was variable among surgeons (κ = 0.16-0.73) and fair between surgeons (κ = 0.23-0.32). The average percentage of agreement for the acetabular component was 47% (range, 32%-64%) and for the femoral component was 54% (range, 38%-70%). Surgeons tended to underestimate implant size if the correct implant was not chosen (acetabular, 29%; femoral, 32%). Selection of an undersized femoral component may lead to femoral neck notching or varus implant alignment. This study emphasizes the need for intraoperative verification of preoperative templating results to ensure optimal implant selection in hip resurfacing.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 25, Issue 3, April 2010, Pages 445–449
نویسندگان
, , , , , ,