کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4062729 1604032 2011 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Posterior Condyle Surface Damage on Retrieved Femoral Knee Components
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Posterior Condyle Surface Damage on Retrieved Femoral Knee Components
چکیده انگلیسی

Twenty-two retrieved femoral knee components were identified with posterior condyle surface damage on average at 99° flexion (range, 43°-135° flexion). Titanium alloy material transfer and abrasive surface damage were evident on cobalt-chromium alloy femoral components that were in contact with titanium alloy tibial trays. Surface damage on the retrieved Oxinium femoral components (Smith and Nephew, Inc, Memphis, Tenn) that were in contact with titanium alloy tibial trays showed gouging, associated with the removal and cracking of the oxide and exposure of the zirconium-niobium alloy substrate. Cobalt-chromium alloy femoral components that were in contact with cobalt-chromium alloy tibial trays showed abrasive wear. Contact between the femoral component and tibial tray should be avoided to prevent surface damage to the femoral condyles, which could potentially accelerate polyethylene wear in vivo.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 26, Issue 8, December 2011, Pages 1460–1467
نویسندگان
, , , ,