کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4062739 1604032 2011 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Radiographic Analysis of a Hand-Held Surgical Navigation System for Tibial Resection in Total Knee Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Radiographic Analysis of a Hand-Held Surgical Navigation System for Tibial Resection in Total Knee Arthroplasty
چکیده انگلیسی

Tibial intramedullary or extramedullary alignment guides have not been shown to be highly accurate in performing the tibial resection in total knee arthroplasty (TKA). Since May 2010, a total of 42 knees underwent a TKA using a hand-held, accelerometer-based surgical navigation system for performing the tibial resection (KneeAlign; OrthAlign Inc, Aliso Viejo, Calif). Postoperative standing anteroposterior hip-to-ankle and lateral knee-to-ankle radiographs demonstrated that 97.6% of the tibial components were placed within 90° ± 2° to the mechanical axis in the coronal plane, and 96.2% of the components were placed within 3° ± 2° to the mechanical axis in the sagittal plane. The KneeAlign greatly improves the accuracy of tibial component alignment in TKA.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 26, Issue 8, December 2011, Pages 1527–1533