کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4062739 | 1604032 | 2011 | 7 صفحه PDF | دانلود رایگان |

Tibial intramedullary or extramedullary alignment guides have not been shown to be highly accurate in performing the tibial resection in total knee arthroplasty (TKA). Since May 2010, a total of 42 knees underwent a TKA using a hand-held, accelerometer-based surgical navigation system for performing the tibial resection (KneeAlign; OrthAlign Inc, Aliso Viejo, Calif). Postoperative standing anteroposterior hip-to-ankle and lateral knee-to-ankle radiographs demonstrated that 97.6% of the tibial components were placed within 90° ± 2° to the mechanical axis in the coronal plane, and 96.2% of the components were placed within 3° ± 2° to the mechanical axis in the sagittal plane. The KneeAlign greatly improves the accuracy of tibial component alignment in TKA.
Journal: The Journal of Arthroplasty - Volume 26, Issue 8, December 2011, Pages 1527–1533