Computed Tomography Measurement of the Accuracy of Component Version in Total Hip Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Computed Tomography Measurement of the Accuracy of Component Version in Total Hip Arthroplasty
چکیده انگلیسی

The version of the acetabular and femoral components in 111 primary total hip arthroplasties was prospectively evaluated intraoperatively by the surgeon and compared with postoperative computed tomography (CT) scan measurements. Intraoperative estimations by the surgeons for acetabular and femoral components were all within 10° to 30° anteversion, with means of 16.0° (SD = 4.0°) and 16.4° (SD = 3.2°), respectively. However, CT scan acetabular measurements ranged from 12° retroversion to 52° anteversion (mean = 22.0° anteversion, SD = 14.0°). Similarly, femoral component version ranged from −15° retroversion to 45° anteversion (mean = 16.8° anteversion, SD = 11.1°). According to CT calculations, only 71% of femoral and 45% of acetabular components were within the expected clinical version range. In conclusion, the intraoperative estimation of acetabular and femoral version in a total hip arthroplasty is of limited accuracy.