کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4062908 | 1604040 | 2011 | 8 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: Cementing Techniques for Hip Resurfacing Arthroplasty: In Vitro Study of Pressure and Temperature Cementing Techniques for Hip Resurfacing Arthroplasty: In Vitro Study of Pressure and Temperature](/preview/png/4062908.png)
Cementing irregularities have been associated with femoral failures of resurfacing arthroplasties in retrieval studies. We used an in vitro model to measure pressure, temperatures, and cement penetration as a function of 6 different cementing techniques. Filling the component with cement can lead to overpenetration or increase the resistance to component seating with resultant polar cement mass. Both conditions result in high and long-lasting cement pressures, cement defects, as well as peak temperatures higher than 50°C. Manual application of cement provides complete penetration of the available fixation area with the lowest cement pressures, the smallest total cement mass, and a peak temperature of 36.0°C ± 4.1°C. Application of the principles elucidated by this study may reduce the risk of cement overpenetration and incomplete seating.
Journal: The Journal of Arthroplasty - Volume 26, Issue 1, January 2011, Pages 144–151