کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4062976 | 1604035 | 2011 | 6 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: Initial Experience of the Journey-Deuce Bicompartmental Knee Prosthesis : A Review of 36 Cases Initial Experience of the Journey-Deuce Bicompartmental Knee Prosthesis : A Review of 36 Cases](/preview/png/4062976.png)
We evaluated the effectiveness of a novel bicompartmental knee arthroplasty prosthesis for the treatment of degenerative disease affecting the medial and patellofemoral compartments. The study included 36 knees in 32 patients with a mean follow-up of 21 months. The mean Knee Society functional survey and Western Ontario McMaster Osteoarthritic Index Survey scores were 65.4 and 75.8, respectively. Thirty-one percent of patients were unsatisfied with the surgery, and 53% stated that they would not repeat the surgery. We report an overall survival rate of 86% with one catastrophically failed tibial baseplate. We conclude that this prosthesis provides inconsistent pain relief and unacceptable functional results for bicompartmental arthritis. Short-term survival of this prosthesis was unacceptably low, and therefore, we no longer implant it at our institution.
Journal: The Journal of Arthroplasty - Volume 26, Issue 6, Supplement, September 2011, Pages 40–45