کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063021 1604089 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Fractures of Cementless Thin-Walled Cups
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Fractures of Cementless Thin-Walled Cups
چکیده انگلیسی

Cementless threaded cups are commonly used in total hip arthroplasty. A method to match the elasticity of the cup to that of natural bone is reducing the wall thickness of the implant. Despite good results with this philosophy of implantation, we recently observed 3 cases of implant fractures after the implantation of such thin-walled cups called “Bicon.” We performed a thorough analysis of 1 of these cases including a histological examination and a technical failure analysis (including scanning electron microscopy) to establish the chronology and cause of the failure. Multiple fractures of the thin-walled metal-back cup were determined to be caused by fatigue. We have concluded that due to the specific material and design parameters of this cup, under adverse circumstances such as the lack of primary or secondary osteointegration, there is a risk of failure due to fatigue.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 21, Issue 1, January 2006, Pages 144–147
نویسندگان
, , , , , ,