کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063119 1604088 2006 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Three-Dimensional Analysis of Alignment Error in Using Femoral Intramedullary Guides in Unicompartmental Knee Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Three-Dimensional Analysis of Alignment Error in Using Femoral Intramedullary Guides in Unicompartmental Knee Arthroplasty
چکیده انگلیسی

We used computerized simulations with 3-dimensional models of 20 cadaver femora, calculated from computed tomographic scans, and a model of a rod measuring 200 × 5 mm to study femoral alignment accuracy for unicompartmental knee arthroplasty via minimally invasive reconstruction. The anatomical axis and insertion site were identified on each femur. A simulation of all feasible flexion-extension and varus-valgus orientations was performed. The average rod orientation was 3.2° flexion and 2.5° valgus. The range of orientation was 3.2° extension to 9.7° flexion and 4.5° varus to 8.9° valgus. The study suggests that a short narrow intramedullary rod inserted according to the manufacturer's specifications does not accurately find the anatomical axis and may lead to poor alignment of the femoral prosthesis. Given our finding of consistent bias toward excessive flexion and valgus alignment, we recommend that the operating surgeon carefully plan the insertion point of the intramedullary rod during surgery to compensate for this bias.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 21, Issue 2, February 2006, Pages 271–278
نویسندگان
, , , ,