کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063207 1604037 2011 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Fatigue Fracture of a Short Stem Hip Replacement: A Failure Analysis With Electron Microscopy and Review of the Literature
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Fatigue Fracture of a Short Stem Hip Replacement: A Failure Analysis With Electron Microscopy and Review of the Literature
چکیده انگلیسی
Four years after primary implantation of an ESKA Cut short-stemmed prosthesis, a 61-year-old patient suffered fracture of the implant. The combination of missing proximal support of the prosthesis as well as enormous strain on the distal third of the stem can be considered as main reason for the fatigue failure. The special design of the modular implant with a small diameter and a spongy metal surface can be regarded as a contributing factor. This unique case demonstrates a possible failure mechanism of short- and ultra short-stemmed prostheses.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 26, Issue 4, June 2011, Pages 665.e17-665.e20
نویسندگان
, , ,