کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063240 1604056 2009 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Measurement of the Acetabular Cup Anteversion on Simulated Radiographs
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Measurement of the Acetabular Cup Anteversion on Simulated Radiographs
چکیده انگلیسی

Widmer (J Arthroplasty 2004;19:387) reported a protractor for measuring the anteversion of acetabular cups on radiographs but with limited precision. We intended to improve its precision by trigonometric mathematics. We measured the anteversion of the acetabular cups on 336 simulated radiographs using aforementioned 2 methods. The anteversion measured by Widmer's protractor ranged from 7° to 41° (mean ± SD = 28.0° ± 9.8°), and our methods, 5° to 51° (27.7° ± 13.2°). The mean ± SD of error by Widmer's protractor was 5.2 ± 2.5°, and our protractor, 0.8° ± 0.8° (Student t test, P < .0001). The interobserver study showed the difference between measurements less than 2° for each method. Therefore, the smaller error of our method than that of Widmer implicated a potentially precise measurement of the anteversion (level of evidence: diagnostic study, level II).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 24, Issue 3, April 2009, Pages 468–474
نویسندگان
, , , , ,