کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4063258 | 1604046 | 2010 | 4 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: The Accuracy of Digital Templating in Uncemented Total Hip Arthroplasty The Accuracy of Digital Templating in Uncemented Total Hip Arthroplasty](/preview/png/4063258.png)
We analyzed the reproducibility and reliability of computer templating in primary uncemented total hip arthroplasties as compared to standard onlay templating techniques with hardcopy radiographs from a digital source. Digital templating showed good intraobserver and interoberserver reliability with intraclass correlation coefficient values greater than 0.7. Using computer templating, prediction of sizing to within 1 size was 85% accurate for femoral stem sizing and 80% accurate for acetabular sizing. Using onlay templating, prediction of sizing to within 1 size was 85% accurate for femoral sizing and 60% accurate for acetabular sizing. We conclude that the introduction of digital templating has significant benefits in preoperative planning for total hip arthroplasty over onlay templating.
Journal: The Journal of Arthroplasty - Volume 25, Issue 4, June 2010, Pages 529–532