کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063271 1604046 2010 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Computer-Assisted Hip Resurfacing Using Individualized Drill Templates
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Computer-Assisted Hip Resurfacing Using Individualized Drill Templates
چکیده انگلیسی

The goal of this study was to investigate whether individualized templates can provide an accurate and reliable computer-assisted system for femoral component placement during hip resurfacing. A consecutive series of 45 patients were examined. Using a 3-dimensional computer model of the femur, the drill trajectory for the central pin of the stem was planned. A surface-matched plastic drilling template was created using a rapid prototyping machine. This patient-specific drill guide was intraoperatively positioned on the patient anatomy, the central pin was drilled into the femoral neck, and the accuracy of the placement with respect to the planned central pin alignment was measured. With mean deviation between planned and actual central pin alignment of 1.14° in varus and 4.49° in retroversion, individualized templates were as accurate as conventional computer-assisted hip resurfacing.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 25, Issue 4, June 2010, Pages 600–606
نویسندگان
, , , ,