کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063421 1604058 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Fracture of a Cross-Linked Polyethylene Liner Due to Impingement
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Fracture of a Cross-Linked Polyethylene Liner Due to Impingement
چکیده انگلیسی
We report a case of fracture at 2 years after implantation of a 50-kGy moderately cross-linked ultrahigh molecular weight polyethylene liner with an extended lip (Marathon, DePuy, Warsaw, IN). The extended lip section had fractured. The liner showed no oxidation. The articular surface was grossly deformed, likely due to wear, creep, and/or plastic deformation, and the liner showed no recovery of machining marks upon melting, indicating that some wear had occurred. Electron microscopy revealed fatigue striations on the fracture surface. The likely cause of failure was femoral neck impingement-induced wear and fatigue on the liner.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 24, Issue 1, January 2009, Pages 158.e15-158.e19
نویسندگان
, , , ,