کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063442 1604044 2010 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Influence of Prosthetic Design on Squeaking After Ceramic-on-Ceramic Total Hip Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Influence of Prosthetic Design on Squeaking After Ceramic-on-Ceramic Total Hip Arthroplasty
چکیده انگلیسی

Squeaky ceramic-on-ceramic (COC) total hips have received much recent publicity, and implant design may be implicated. We reviewed 270 consecutive COC total hip arthroplasties in 233 patients comparing 4 implant combinations representing 4 manufacturers. A cohort (n = 45) of Stryker Trident acetabular cups paired with Stryker Accolade femoral stems showed a dramatically higher incidence of “problem squeaking”—defined as always audible to others and occurring at least once per week—with a 35.6% incidence of squeaking and 11.1% incidence of problem squeaking. The 3 non-Stryker designs (n = 225) revealed 3.6% squeaking (P < .0001) and 0.44% problem squeaking (P = .006). The Stryker system has a unique design and metallurgy. Our results suggest that although the genesis of squeaking in COC total hips is multifactorial, prosthetic design plays a key role.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 25, Issue 6, Supplement, September 2010, Pages 36–42