Steady-State Penetration Rates of Electron Beam–Irradiated, Highly Cross-Linked Polyethylene at an Average 45-Month Follow-Up
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Steady-State Penetration Rates of Electron Beam–Irradiated, Highly Cross-Linked Polyethylene at an Average 45-Month Follow-Up
چکیده انگلیسی

Steady-state penetration rates were determined on 53 total hip arthroplasties with a minimum 3-year follow-up using electron beam highly cross-linked melted polyethylene by determining the femoral head penetration occurring after the first year in vivo. These data were compared with that of a matched control group. The average steady-state penetration rate of the control group was 144 ± 191 μm/y, significantly higher than that of the highly cross-linked polyethylene group, 25 ± 99 μm/y (P = .0001). Linear regression analysis indicated that the magnitude of femoral head penetration did not increase with time in the cross-linked groups. Also, no significant difference existed between the steady-state wear rates of the highly cross-linked groups with two head sizes (28 vs 32 mm, P = .39).