کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063995 1604081 2006 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Influence of Polyethylene Constraint on Tibial Component Fixation in Total Knee Arthroplasty. Follow-Up Report After 5 Years
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Influence of Polyethylene Constraint on Tibial Component Fixation in Total Knee Arthroplasty. Follow-Up Report After 5 Years
چکیده انگلیسی

The influence of articulating surface design of AMK total knee prosthesis (DePuy, Johnson & Johnson) on migration and radiographic outcome at 5 years was evaluated. The knees were randomly allocated to receive a flat or a concave insert with retention of the posterior cruciate ligament when preoperative deformity was less severe and either a concave or a posterior-stabilized insert with resection of the posterior cruciate ligament when deformity was more pronounced. In 64 knees, migration was measured with radiostereometry. The posterior-stabilized component displayed more varus-valgus tilting than the concave insert. Other statistically significant differences in migration were not seen. Radiolucent lines were frequently seen without differences between prosthesis groups.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 21, Issue 7, October 2006, Pages 1032–1037
نویسندگان
, , , , ,