کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4066813 1604357 2015 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
The Effect of Barbed Suture Tendon Repair on Work of Flexion
ترجمه فارسی عنوان
تأثیر اصلاح طناب خیاطی در کار فشرده سازی
کلمات کلیدی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی

PurposeTo compare the work of flexion, ultimate strength, and gap resistance of a conventional 4-strand tendon repair to a knotless barbed-suture 4-strand tendon repair.MethodsTendon repairs were performed on 16 cadaver flexor digitorum profundus tendons using either a 4-strand double Kessler repair or a similar but knotless 4-strand repair with a unidirectional barbed suture. Work of flexion, gap resistance during cyclical loading, and ultimate strength of both techniques were determined and their means compared.ResultsThere was no difference in mean maximum load and gap formation between the 2 techniques. Work of flexion was higher for the barbed-suture repair group compared with the traditional repair group (39 N·mm vs 31 N·mm).ConclusionsThe higher work of flexion in the barbed-suture group suggests that barbed suture may negatively affect tendon gliding within the flexor tendon sheath.Clinical relevanceKnotless barbed-suture tendon repair leads to increased work of flexion compared with traditional flexor tendon repairs, which may result in an increased rupture incidence.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Hand Surgery - Volume 40, Issue 5, May 2015, Pages 969–974
نویسندگان
, , , ,