کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4074733 1267019 2008 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of conforming and nonconforming retrieved glenoid components
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Comparison of conforming and nonconforming retrieved glenoid components
چکیده انگلیسی

The purpose of this study was to compare differences in wear performance of conforming and nonconforming glenoid designs, using clinical, radiographic, and retrieved polyethylene glenoid component outcome Sixty-three glenoids met the study criteria, and each glenoid was assigned to the conforming group (if the radii of curvature of the humeral and glenoid components were identical) or the nonconforming group (if a mismatch existed between the radii of curvature). The mean length of implantation was 5.6 ± 5.5 years in the conforming group and 3.1 ± 3.1 years for the nonconforming group (P < .05). The loosening score was 3.2 ± 2.0 in the conforming group and 2.4 ± 1.2 in the nonconforming one (P < .05). The nonconforming group had a significantly greater burnishing score (P < .01), while the conforming group had greater abrasion and delamination scores (P < .05). Articular conformity contributes to differences observed from retrieved polyethylene glenoid components, which are consistent with differences in performance that may influence loosening.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Shoulder and Elbow Surgery - Volume 17, Issue 6, November–December 2008, Pages 914–920