کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4074985 1267027 2007 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Biomechanical comparison of component position and hardware failure in the reverse shoulder prosthesis
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Biomechanical comparison of component position and hardware failure in the reverse shoulder prosthesis
چکیده انگلیسی

There has been renewed interest in reverse shoulder arthroplasty for the treatment of glenohumeral arthritis with concomitant rotator cuff deficiency. Failure of the prosthesis at the glenoid attachment site remains a concern. The purpose of this study was to examine glenoid component stability with regard to the angle of implantation. This investigation entailed a biomechanical analysis to evaluate forces and micromotion in glenoid components attached to 12 polyurethane blocks at −15°, 0°, and +15° of superior and inferior tilt. The 15° inferior tilt had the most uniform compressive forces and the least amount of tensile forces and micromotion when compared with the 0° and 15° superiorly tilted baseplate. Our results suggest that implantation with an inferior tilt will reduce the incidence of mechanical failure of the glenoid component in a reverse shoulder prosthesis.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Shoulder and Elbow Surgery - Volume 16, Issue 3, Supplement, May–June 2007, Pages S9–S12
نویسندگان
, , , , ,