In vitro and finite element analysis of glenoid bone/baseplate interaction in the reverse shoulder design
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
In vitro and finite element analysis of glenoid bone/baseplate interaction in the reverse shoulder design
چکیده انگلیسی

We developed biomechanical and finite element models, using high-strength polyurethane foam blocks, to represent the glenoid bone/baseplate junction to determine if increasing the distance between the glenoid bone and the center of rotation of the glenosphere increases baseplate motion during static loading in the reverse shoulder design. Although there was a general trend toward increased baseplate motion with increasing distance from the glenoid to the center of rotation, in vitro mechanical testing revealed no significant difference between the 7 glenosphere types tested, with average baseplate motion during 1000 load cycles ranging from 90 μm to 120 μm. Results from the finite element analysis strongly correlated with the in vitro mechanical testing. The magnitude of baseplate motion occurring in a modeled representation of bone under simulated physiologic loading conditions was similar for the 7 reverse shoulder glenoid components tested in this study.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Shoulder and Elbow Surgery - Volume 17, Issue 3, May–June 2008, Pages 509–521