Intraoperative assessment of midflexion laxity in total knee prosthesis
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Intraoperative assessment of midflexion laxity in total knee prosthesis
چکیده انگلیسی


• We developed a new tensor device, which could attach polyethylene insert trial.
• We measured the joint gap in 259 knees throughout full range of motion.
• The size of the joint gap was loose at midflexion ranges (p < 0.001).
• The symmetry of the joint gap was varus at 0° and 145° of flexion (p < 0.001).

BackgroundSoft-tissue balancing of the knee is fundamental to the success of a total knee arthroplasty (TKA). In posterior-stabilized TKA, there is no stabilizer of the anterior–posterior translation in the midflexion range in which the cam-post mechanism does not engage yet. Therefore, instability in the midflexion range is suspected to occur in posterior-stabilized TKA. The purpose of this study was to measure the joint gap throughout a full range of motion and to analyze the joint gap laxity in the midflexion range after implantation of a mobile-bearing posterior-stabilized total knee prosthesis.MethodsJoint gap kinematics in 259 knees with varus osteoarthritis were measured during TKAs using a tensor device with the same shape of a total knee prosthesis of the same design was used. After the implantation of a mobile-bearing posterior-stabilized prosthesis and the reduction of the patellofemoral joint, the joint gap was measured at 0°, 30°, 60°, 90°, 120°, and 145° of flexion.ResultsThe center size of the joint gap was tight in extension and deep flexion and loose at midflexion ranges, especially at 30° of flexion (p < 0.001). The symmetry of the joint gap was varus at 0° and 145° of flexion (p < 0.001).ConclusionsOur results showed the joint gap laxity in the midflexion range after the implantation of a mobile-bearing posterior-stabilized prosthesis. Our new tensor device, which can attach the polyethylene insert trial, will provide the important information about the joint gap kinematics after implantation of total knee prostheses.Level of evidenceIV.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Knee - Volume 21, Issue 4, August 2014, Pages 810–814