Explant Analysis of Total Disc Replacement
چکیده انگلیسی
Explant analysis of human disc prostheses allow early evaluation of the host response to the prosthesis and the response of the prosthesis from the host. Furthermore, early predictions of failure and wear can be obtained. Thus far, about 2-3% of disc prostheses have been removed. Observed wear patterns are similar to that of appendicular prostheses including abrasions/scratching, burnishing, surface deformation, fatigue, and embedded debris. Chemically the polymeric components have shown little degradation in short-term implantation. In metal on metal prostheses the histologic responses consist of large numbers of metallic particles with occasional macrophages and giant cells. Only rare cases of significant inflammatory response from polymeric debris have been seen.