کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4106467 1605390 2011 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison: real and simulated ear insertion gain
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیماری های گوش و جراحی پلاستیک صورت
پیش نمایش صفحه اول مقاله
Comparison: real and simulated ear insertion gain
چکیده انگلیسی

The development of hearing aid (HA) software programming does not replace the analysis of real ears with probe microphones.AimTo compare simulated insertion gain in HA software programming and real ear insertion gain.MethodA prospective study of 62 patients (aged from 29 to 93 years; 30 male and 32 female). All patients presented unilateral (n=14) or bilateral (n=48) and mild to profound sensorineural hearing impairment. 110 ears assessed. Data was gathered from medical records and the insertion gain was obtained in real ears for comparison with the simulated insertion gain in HA software programming. Statistical tests were applied to analyze the correlation of data - difference of real ear and simulated insertion gain.ResultsHA software programming simulated insertion gain was higher than real ear insertion gain obtained with probe microphone measures. There were statistically significant frequency differences. Age did not correlate with the difference of real ear and simulated measures.ConclusionThe use of real ear measures is important during verification of HA.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Brazilian Journal of Otorhinolaryngology - Volume 77, Issue 5, September–October 2011, Pages 555–558