Accuracy of Pulse Oximetry Measurement of Heart Rate of Newborn Infants in the Delivery Room
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی پریناتولوژی (پزشکی مادر و جنین)، طب اطفال و بهداشت کودک
پیش نمایش صفحه اول مقاله
Accuracy of Pulse Oximetry Measurement of Heart Rate of Newborn Infants in the Delivery Room
چکیده انگلیسی

ObjectiveTo determine the accuracy of heart rate obtained by pulse oximetry (HRPO) relative to HR obtained by 3-lead electrocardiography (HRECG) in newborn infants in the delivery room.Study designImmediately after birth, a preductal PO sensor and ECG leads were applied. PO and ECG monitor displays were recorded by a video camera. Two investigators reviewed the videos. Every two seconds, 1 of the investigators recorded HRPO and indicators of signal quality from the oximeter while masked to ECG, whereas the other recorded HRECG and ECG signal quality while masked to PO. HRPO and HRECG measurements were compared using Bland-Altman analysis.ResultsWe attended 92 deliveries; 37 infants were excluded due to equipment malfunction. The 55 infants studied had a mean (±standard deviation [SD]) gestational age of 35 (±3.7) weeks, and birth weight 2399 (±869) g. In total, we analyzed 5877 data pairs. The mean difference (±2 SD) between HRECG and HRPO was −2 (±26) beats per minute (bpm) overall and −0.5 (±16) bpm in those infants who received positive-pressure ventilation and/or cardiac massage. The sensitivity and specificity of PO for detecting HRECG <100 bpm was 89% and 99%, respectively.ConclusionPO provided an accurate display of newborn infants' HR in the delivery room, including those infants receiving advanced resuscitation.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Pediatrics - Volume 152, Issue 6, June 2008, Pages 756–760
نویسندگان
, , , , , , ,