The accuracy of three-dimensional bladder ultrasonography in determining the residual urinary volume compared with conventional catheterisation
ترجمه فارسی عنوان
دقت سونوگرافی سه بعدی مثانه در تعیین حجم ادرار باقیمانده در مقایسه با سونوگرافی معمولی
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی اورولوژی
چکیده انگلیسی

ObjectiveTo determine the accuracy of three-dimensional bladder ultrasonography (US, using the BVI 3000, Verathon, WA, USA) for determining the residual urinary volume, compared with the conventional catheterisation method.Patients and methodsWe conducted a cross-sectional study at day-care unit of a University hospital after obtaining approval from the Ethics Review Committee of the hospital. Thirty-four patients with lower urinary tract symptoms requiring cystometrography were included. The postvoid volume was measured by bladder US, with three readings taken, and then patient was catheterised using a 12-F Nelaton catheter to measure the urinary volume. The mean of the three readings was compared with the catheterisation volume.ResultsThe mean (SD) urinary volumes by US and catheterisation were 261 (186) and 260 (175) mL, respectively, and the correlation (r2) was 0.97. There was no effect of age, gender or body mass index on the accuracy of bladder US, which was accurate even when the urinary volume was ⩽100 mL.ConclusionThe bladder US estimate is as accurate as catheterisation for determining the postvoid residual urinary volume. Its accuracy was also comparable when the urinary volume is <100 mL, and there was no significant effect of age, gender and body mass index. This system could replace the more invasive catheterisation, and with excellent accuracy.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Arab Journal of Urology - Volume 12, Issue 3, September 2014, Pages 209–213
نویسندگان
, ,