BackgroundRadiofrequency ablation is a popular method of treating unresectable liver tumors but tumors greater than 3 cm in diameter have a much greater risk of local recurrence after treatment. Bimodal electric tissue ablation is a modified form of radiofrequency ablation that creates significantly larger ablations by the addition of extra direct current circuitry. This may help to reduce the risk of local recurrence in these larger tumors. Prior to use in a clinical setting, a long term study was performed to assess associated morbidity and the pathological changes in the ablations.MethodsIn eight pigs, six ablations were performed in each liver. Pigs were euthanized at 2 d, 2 wk, 2 mo, and 4 mo, and the ablations were assessed macroscopically and microscopically for pathological change. Regular blood tests were performed to assess changes in liver function. At death, any other abnormalities detected were reported.ResultsHistopathological examination of ablation zones revealed tissue death by coagulative necrosis and healing by fibrotic scarring. Transient rises in serum liver enzymes were seen in the postoperative period. Skin necrosis was noted at the site of the positive electrode of the direct electrical current but no other form of morbidity was seen associated with the procedure.ConclusionsAlthough the positive electrode placement requires further consideration, bimodal electric tissue ablation appears to be safe and behaves in a similar fashion to other thermal therapies such as standard radiofrequency ablation.
Journal: Journal of Surgical Research - Volume 148, Issue 2, August 2008, Pages 251–259